- Ideal for outdoor as well as panel mount application
- Conductivity/TDS meter with inbuilt relay & ALARM output with acknowledge facility
- Programmable hysteresis avoids unnecessary chattering of Pump/Solenoid Valve etc
- Programmable control delay time
- Automatic temperature compensation (optionally available)
- 4-20 mA output for panel indication/PLC (optionally available)
- Communication ready for monitoring/data logging (optionally available)
[/bsf-info-box]
[bsf-info-box icon=”Defaults-bookmark” icon_size=”18″ icon_color=”#1e73be” icon_style=”advanced” icon_border_style=”solid” icon_color_border=”#1e73be” icon_border_size=”2″ icon_border_radius=”500″ icon_border_spacing=”30″ title=”Available Ranges “]0-19.99 µs/cm²
0-1999 µs/cm²
0-9999 µs/cm²[/bsf-info-box]
[bsf-info-box icon=”Defaults-bookmark” icon_size=”18″ icon_color=”#1e73be” icon_style=”advanced” icon_border_style=”solid” icon_color_border=”#1e73be” icon_border_size=”2″ icon_border_radius=”500″ icon_border_spacing=”30″ title=”Specifications”]
Physical Dimensions | 105 x 105 x 130 mm |
Cut out Size | 92*92 mm |
Enclosure | ABS weather proof IP 65 |
Mounting | Field / Panel |
Input Supply | 230V A.C, 50/60 Hz or 110 V A.C. or 24 V D.C. |
Alarm | In built buzzer with blinking LED for High Cond./TDS indication |
Relay | 1 no. for High(programmable through entire range with Control delay & Hysteresis) 5A @ 230V A.C |
Control Operation | Alarm / Auto reset |
Accuracy | ± 2% of FSD |
Resolution | 0.001 / 0.01 / 0.1 / 1 depending on range |
Display | 4 digit 7 segment LED
OR 16 X 2 alphanumeric backlit LCD |
Calibration / Set Point | Using front panel keys |
Terminals | Screwed on rear side |
Measuring Range | 0-1999 (other ranges available) |
Display mode | µS/cm² or ppm |
Output | 4-20 mA (optional) / RS 485 (optional) |
[/bsf-info-box]
[bsf-info-box icon=”Defaults-bookmark” icon_size=”18″ icon_color=”#1e73be” icon_style=”advanced” icon_border_style=”solid” icon_color_border=”#1e73be” icon_border_size=”2″ icon_border_radius=”500″ icon_border_spacing=”30″ title=”Product Details”]
Conductivity Meter is the direct measure of ionic concentration. The pure the water the lesser the conductivity. Salts, minerals and even dissolved gases contribute uniformly to the conductivity of a solution. This means that the conductivity can be used as an indicator of the amount of dissolved materials in a solution.
[/bsf-info-box]